With pooled sensitivity of 0.77, false-negative rate with pulse CO oximeters too high for evaluating individuals suspected of having CO poisoning
THURSDAY, Oct. 20, 2022 (HealthDay News) — The use of noninvasive pulse carbon monoxide (CO) oximeters (SpCO) is not accurate for determining blood carboxyhemoglobin (COHb) levels, according to a study presented Oct. 18 at the European Emergency Medicine Congress, in Germany.
Mathilde Papin, from Nantes University Hospital in France, and colleagues conducted a meta-analysis to determine the accuracy of SpCO in determining blood COHb level. The systematic review included 19 studies; 11 and 13 were included for quantitative analysis of the primary and secondary end points, respectively.
Three different devices were used to measure SpCO. The principal source of bias and concern regarding applicability was patient selection. There was no publication bias. The researchers found that the area under the summary receiver operating characteristic curve was 86 percent, with pooled sensitivity and specificity of 0.77 and 0.83, respectively. The mean bias was 0.95 percent, and limits of agreement were −6.08 to 8.00 percent.
“At 23 percent, the false negative rate with pulse oximetry is too high for reliably triaging patients with suspected carbon monoxide poisoning. This method is not accurate enough and should not be used in clinical practice,” the authors write.
AAP Urges Regular Eye Checks for Children With JIA, Hydroxychloroquine Users
One-Hour Cognitive Test Provides Meaningful Information Beyond Traditional Memory Screening
Broad Genomic Sequencing Improves Survival in First-Line Setting in Advanced NSCLC
Screening, Treating STIs During Pregnancy May Cut Preterm Birth Risk
Early Postpartum Brain Loss Prominent but Reversible in MS
Prevalence of Preeclampsia Increased With Familial Hypercholesterolemia
Self-Led, Digital Intervention Lowers Anxiety During Pregnancy
Guidelines Updated for Prostate Cancer Screening